Indexed by:
Abstract:
Glucose oxidase(GOD) was encapsulated in the Graphene/Nafion film modified glassy carbon electrode(GCE) and used as an ECL sensor for glucose. The GOD retains its bioactivity after being immobilized into the composite film. The sensor gives a linear response for glucose in the range of 2.0 x 10(-6)-1.0 x 10(-4) mol/L with a detection limit of 1.0 x 10(-6) mol/L. The sensor showed good stability, the RSD for continuous scanning for 5.0 x 10(-5) mol/L glucose was 4.21% (n = 5). After being stored in 0.05 mol/L pH 7.4 PBS in 4 degrees C for two weeks, the modified electrode maintains 80% of its initial activity. The glucose sensor provides new opportunity for clinical diagnosis applications.
Keyword:
Reprint 's Address:
Email:
Version:
Source :
ELECTROANALYSIS
ISSN: 1040-0397
Year: 2010
Issue: 20
Volume: 22
Page: 2347-2352
2 . 7 2 1
JCR@2010
2 . 7 0 0
JCR@2023
ESI Discipline: CHEMISTRY;
JCR Journal Grade:2
CAS Journal Grade:3
Cited Count:
WoS CC Cited Count: 49
SCOPUS Cited Count: 52
ESI Highly Cited Papers on the List: 0 Unfold All
WanFang Cited Count:
Chinese Cited Count:
30 Days PV: 0
Affiliated Colleges: