Abstract:
Diabetes is an important public health problem and finding quick testing methods with high accuracy, reliability, and convenience are important to control the blood glucose of diabetic patients. In this study, a sensor based on a weak measurement scheme was developed for the specific detection of glucose for the first time. The detection of glucose using the proposed method was completed by the high sensitivity and resolution of the weak measurement based on optical rotation detection, as well as the change in the optical rotation before and after the specific oxidation of glucose. The resolution of the as-obtained glucose sensor was around 2.71×10-3 g/L (1.50×10-2 mmol/L), and the detection range was 0-11 g/L (0-61 mmol/L). © 2021 Optical Society of America.