Indexed by:
Abstract:
详细分析了P47型原子力显微镜线宽测量不确定度的来源,给出了基于几何形状的线宽测量模型,提出了线宽测量不确定度的评定路线和方法.确定了探针针尖引起的测量不确定度是AFM线宽测量不确定度的主要来源,并对其进行了定量分析.普通Si3N4探针针尖引起的不确定度分量约占线宽总量的5%.
Keyword:
Reprint 's Address:
Email:
Version:
Source :
计量学报
ISSN: 1000-1158
CN: 11-1864/TB
Year: 2012
Issue: 6
Volume: 33
Page: 486-489
Cited Count:
SCOPUS Cited Count: 1
ESI Highly Cited Papers on the List: 0 Unfold All
WanFang Cited Count: -1
Chinese Cited Count:
30 Days PV: 0