The aim of this study was to verify the feasibility and accuracy of a contour registration-based augmented reality (AR) system in jaw surgery. An AR system was developed to display the interaction between virtual planning and images of the surgical site in real time. Several trials were performed with the guidance of the AR system and the surgical guide. The postoperative cone beam CT (CBCT) data were matched with the preoperatively planned data to evaluate the accuracy of the system by comparing the deviations in distance and angle. All procedures were performed successfully. In nine model trials, distance and angular deviations for the mandible, reconstructed fibula, and fixation screws were 1.62 ± 0.38 mm, 1.86 ± 0.43 mm, 1.67 ± 0.70 mm, and 3.68 ± 0.71°, 5.48 ± 2.06°, 7.50 ± 1.39°, respectively. In twelve animal trials, results of the AR system were compared with the surgical guide. Distance deviations for the bilateral condylar outer poles were 0.93 ± 0.63 mm and 0.81 ± 0.30 mm, respectively (p = 0.68). Distance deviations for the bilateral mandibular posterior angles were 2.01 ± 2.49 mm and 2.89 ± 1.83 mm, respectively (p = 0.50). Distance and angular deviations for the mandible were 1.41 ± 0.61 mm, 1.21 ± 0.18 mm (p = 0.45), and 6.81 ± 2.21°, 6.11 ± 2.93° (p = 0.65), respectively. Distance and angular deviations for the reconstructed tibiofibular bones were 0.88 ± 0.22 mm, 0.84 ± 0.18 mm (p = 0.70), and 6.47 ± 3.03°, 6.90 ± 4.01° (p = 0.84), respectively. This study proposed a contour registration-based AR system to assist surgeons in intuitively observing the surgical plan intraoperatively. The trial results indicated that this system had similar accuracy to the surgical guide. © 2023 The Authors

Keyword:

Augmented reality Mandibular surgery Markerless registration Surgery simulation Surgical guide

Community:

  • [ 1 ] [Liu S.]Department of Oral and Maxillofacial Surgery, First Affiliated Hospital of Fujian Medical University, Fuzhou, 350005, China
  • [ 2 ] [Liu S.]School and Hospital of Stomatology, Fujian Medical University, Fuzhou, 350004, China
  • [ 3 ] [Liao Y.]Department of Oral and Maxillofacial Surgery, First Affiliated Hospital of Fujian Medical University, Fuzhou, 350005, China
  • [ 4 ] [Liao Y.]Laboratory of Facial Plastic and Reconstruction, Fujian Medical University, Fuzhou, 350004, China
  • [ 5 ] [He B.]School of Mechanical Engineering and Automation, Fuzhou University, Fuzhou, 350108, China
  • [ 6 ] [He B.]Fujian Engineering Research Center of Joint Intelligent Medical Engineering, Fuzhou, 350108, China
  • [ 7 ] [Dai B.]Department of Oral and Maxillofacial Surgery, Second Xiangya Hospital of Central South University, Changsha, 410000, China
  • [ 8 ] [Zhu Z.]School of Mechanical Engineering and Automation, Fuzhou University, Fuzhou, 350108, China
  • [ 9 ] [Zhu Z.]Fujian Engineering Research Center of Joint Intelligent Medical Engineering, Fuzhou, 350108, China
  • [ 10 ] [Shi J.]School of Mechanical Engineering and Automation, Fuzhou University, Fuzhou, 350108, China
  • [ 11 ] [Shi J.]Fujian Engineering Research Center of Joint Intelligent Medical Engineering, Fuzhou, 350108, China
  • [ 12 ] [Huang Y.]Department of Oral and Maxillofacial Surgery, First Affiliated Hospital of Fujian Medical University, Fuzhou, 350005, China
  • [ 13 ] [Huang Y.]Laboratory of Facial Plastic and Reconstruction, Fujian Medical University, Fuzhou, 350004, China
  • [ 14 ] [Zou G.]Department of Oral and Maxillofacial Surgery, First Affiliated Hospital of Fujian Medical University, Fuzhou, 350005, China
  • [ 15 ] [Zou G.]Laboratory of Facial Plastic and Reconstruction, Fujian Medical University, Fuzhou, 350004, China
  • [ 16 ] [Du C.]Department of Oral and Maxillofacial Surgery, First Affiliated Hospital of Fujian Medical University, Fuzhou, 350005, China
  • [ 17 ] [Du C.]School and Hospital of Stomatology, Fujian Medical University, Fuzhou, 350004, China
  • [ 18 ] [Shi B.]Department of Oral and Maxillofacial Surgery, First Affiliated Hospital of Fujian Medical University, Fuzhou, 350005, China
  • [ 19 ] [Shi B.]Laboratory of Facial Plastic and Reconstruction, Fujian Medical University, Fuzhou, 350004, China

Reprint 's Address:

Email:

Show more details

Related Keywords:

Related Article:

Source :

Journal of Cranio-Maxillofacial Surgery

ISSN: 1010-5182

Year: 2023

Issue: 6

Volume: 51

Page: 360-368

2 . 1

JCR@2023

2 . 1 0 0

JCR@2023

ESI HC Threshold:25

JCR Journal Grade:2

CAS Journal Grade:1

Cited Count:

WoS CC Cited Count: 0

SCOPUS Cited Count: 1

ESI Highly Cited Papers on the List: 0 Unfold All

WanFang Cited Count:

Chinese Cited Count:

30 Days PV: 0

Affiliated Colleges:

Online/Total:41/10049721
Address:FZU Library(No.2 Xuyuan Road, Fuzhou, Fujian, PRC Post Code:350116) Contact Us:0591-22865326
Copyright:FZU Library Technical Support:Beijing Aegean Software Co., Ltd. 闽ICP备05005463号-1