Abstract:

Hot electron induced cathodic electrochemiluminescence (ECL) was found at a disposable CdS modified screen printed carbon electrodes (CdS-SPCEs) during cathodic pulse polarization. The ECL behavior of the modified electrode was investigated with two emissions located at 520 nm and 580 nm. The CdS-SPCEs show very stable and reproducible ECL signal. The ECL intensity was linear with the dissolved oxygen concentration in the range of 1.7-33 mg/L with a detection limit of 0.02 mg/L (S/N = 3). In addition, a possible mechanism for ECL behavior of the modified electrode was proposed. The developed method can be applied to detect the dissolved oxygen concentration or biochemical oxygen demand (BOD). © 2011 Elsevier B.V.

Keyword:

Cathodic electrochemiluminescence; CdS nanoparticles; Dissolved oxygen; Screen printed carbon electrodes

Community:

  • [ 1 ] [Zheng, R.-J.]Key Laboratory of Analysis and Detection for Food Safety (Ministry of Education), Department of Chemistry, Fuzhou University, Fuzhou 350108, China
  • [ 2 ] [Fang, Y.-M.]Key Laboratory of Analysis and Detection for Food Safety (Ministry of Education), Department of Chemistry, Fuzhou University, Fuzhou 350108, China
  • [ 3 ] [Qin, S.-F.]Key Laboratory of Analysis and Detection for Food Safety (Ministry of Education), Department of Chemistry, Fuzhou University, Fuzhou 350108, China
  • [ 4 ] [Song, J.]Key Laboratory of Analysis and Detection for Food Safety (Ministry of Education), Department of Chemistry, Fuzhou University, Fuzhou 350108, China
  • [ 5 ] [Wu, A.-H.]Key Laboratory of Analysis and Detection for Food Safety (Ministry of Education), Department of Chemistry, Fuzhou University, Fuzhou 350108, China
  • [ 6 ] [Sun, J.-J.]Key Laboratory of Analysis and Detection for Food Safety (Ministry of Education), Department of Chemistry, Fuzhou University, Fuzhou 350108, China

Reprint 's Address:

  • [Sun, J.-J.]Key Laboratory of Analysis and Detection for Food Safety (Ministry of Education), Department of Chemistry, Fuzhou University, Fuzhou 350108, China

Show more details

Source :

Sensors and Actuators, B: Chemical

ISSN: 0925-4005

Year: 2011

Issue: 2

Volume: 157

Page: 488-493

3 . 8 9 8

JCR@2011

6 . 3 9 3

JCR@2018

JCR Journal Grade:1

Cited Count:

WoS CC Cited Count:

SCOPUS Cited Count: 49

30 Days PV: 3

Affiliated Colleges:

Online/Total:106/10140234
Address:FZU Library(No.2 Xuyuan Road, Fuzhou, Fujian, PRC Post Code:350116) Contact Us:0591-22865326
Copyright:FZU Library Technical Support:Beijing Aegean Software Co., Ltd. 闽ICP备05005463号-1