A method was established to determine the activity of hexokinase by reverse-phase high-performance liquid chromatography (RP-HPLC). The activity of hexokinase was determined by detecting the amount of ADP in the reaction that converts glucose into glucose-6-phosphate. The separation degree of ATP and ADP was as good as 3.46 (separation degree >1.5). In the range of 0.01-0. 40 mmol/L, there was a good linear relationship between concentration of ADP and its peak areas (the correlation coefficient was 0.999). The relative standard deviation(RSD) of intraday was 1.43%-1. 46%, that of interday was 2.12%-2.15%. At 30 °C (optimal temperature) and pH7.0 (optimal pH), the recovery rate of hexokinase was 95.3%-97.6%. The results showed that the method had good precision and high recovery rate, and the enzyme activity was determined through only one-step reaction. To some extent, this method can reduce the cost. © 2011 Wuhan University and Springer-Verlag Berlin Heidelberg.