Indexed by:
Abstract:
In this study, we reported on a low detection limit penicillin biosensor with layer-by-layer (LbL) film containing single-graphene nanosheets (SGNs) preadsorbed with hematein, ionic liquids (ILs) and penicillinase. The penicillinase catalyzes the hydrolysis of penicillin to penicilloic acid, where H+ is liberated and monitored amperometrically with hematein as a pH indicator. The SGN-hematein/ILs/penicillinase biosensor exhibited excellent performance for penicillin in PBS with a wide range from 1.25 × 10 - 13 to 7.5 × 10- 3 M, and a low detection limit of 10- 13 M (0.04 ppt, S/N ≥ 3). Furthermore, the detection of penicillin concentration in real sample (milk) had acceptable accuracy with the assay system. © 2014 Elsevier B.V.
Keyword:
Reprint 's Address:
Email:
Source :
Materials Science and Engineering C
ISSN: 0928-4931
Year: 2014
Issue: 1
Volume: 39
Page: 92-99
3 . 0 8 8
JCR@2014
8 . 1 0 0
JCR@2023
JCR Journal Grade:3
CAS Journal Grade:2
Cited Count:
WoS CC Cited Count: 0
SCOPUS Cited Count: 47
ESI Highly Cited Papers on the List: 0 Unfold All
WanFang Cited Count:
Chinese Cited Count:
30 Days PV: 2
Affiliated Colleges: