Indexed by:
Abstract:
A sensitive and rapid method was developed for angiotensin-converting enzyme (ACE) activity determination by micellar electrokinetic capillary chromatography (MECC). MECC was carried out to separate and quantify the products of the enzymatic reacting using Hip-Leu-His as the substrate in 20 mmol/L boric acid-borate buffer (pH 9.0) including 50 mmol/L SDS as the run buffer at an applied voltage of 8.1 kV. The electrophoresis was monitored at 228 nm, and completed in 6 minutes. The detection limits of ACE activity was 5 pmol/min(signal to noise ratio was 2).
Keyword:
Reprint 's Address:
Email:
Source :
Zhongguo hua xue hui
ISSN: 1000-8713
CN: 21-1185/O6
Year: 2001
Issue: 1
Volume: 19
Page: 68-70