Indexed by:
Abstract:
A simple and convenient end-column chemiluminescence (CL) detection coupled to pressurized capillary electrochromatography (pCEC) was described. Luminol and N-(4-aminobutyl)-N-ethylisoluminol (ABEI) were adopted as mode compounds to evaluate the feasibility of end-column reactor. Detailed analysis of ABEI revealed that the high sensitivity could be obtained with the reactor. Furthermore, determination of norepinephrine (NE) and epinephrine (EP), which were labeled with ABEI, was accomplished by using the end-column pCEC-CL detection based on ABEI-potassium ferricyanide-alkaline medium CL reaction system. Under the optimum conditions, the detection limit (S/N = 3) of NE and EP was 0.08 μM and 0.06 μM, respectively. The proposed method has also been successfully applied to the analysis of adrenaline hydrochloride injection sample. © 2007 Elsevier B.V. All rights reserved.
Keyword:
Reprint 's Address:
Email:
Source :
Journal of Chromatography A
ISSN: 0021-9673
Year: 2007
Issue: 1-2
Volume: 1170
Page: 118-121
3 . 6 4 1
JCR@2007
3 . 8 0 0
JCR@2023
JCR Journal Grade:1
Cited Count:
SCOPUS Cited Count: 52
ESI Highly Cited Papers on the List: 0 Unfold All
WanFang Cited Count:
Chinese Cited Count:
30 Days PV: 0
Affiliated Colleges: