Traditional PCR instrument detects the temperature of reagent with contact measurement, which has several limitations such as thermal hysteresis, reagent contamination and point-measurement. Thus an infra red thermometer is applied to contactlessly measure the temperature of reagent. But infra red thermometer has the defect of relatively high measurement noise. Therefore, we propose a sequential dual Kalman filter (Seq-DKF) algorithm that uses an iterated extended Kalman filter (IEKF) and a linear Kalman filter (KF) to remove the measurement noise from infra red thermometer signals in real-time. In this algorithm, IEKF and KF are executed sequentially. The former filter rapidly identifies the parameter of heat exchange model with a small number of observations. The latter one filters the temperature signal based on the identified model. After specifying the heat exchange model and the Seq-DKF, we apply the algorithm to the PCR instrument for testing. Finally, the other three filters are employed for comparison to demonstrate the effectiveness of the Seq-DKF algorithm.