We present the integration and accuracy and power improved hearing aid system that proposes the triple oscillation-based loops (TOL) technique. Compared with the conventional techniques, the technique replaces the interface capacitors, A/D and D/A convertors with the oscillation based closed loop and the calibration/operation feedback paths, which are able to enhance the integration of the system and improve the system's accuracy and power performance. Specifically, the technique achieves the only once of the second-order delta-sigma modulation with an integrator and an oscillator. To realize the architecture, the current mode blocks with calibration and successive subtraction are presented on the circuit design. Simulated with a 0.13 mu m standard CMOS process in Cadence, with the 16 ohm loudspeaker impedance under the 1V supply voltage, the input referred DC o ff set voltage with interface capacitor free is achieved below 2.5 mV; and the signal-to-noise and distortion rate (SNDR) achieves 50 dB@ 400mVp-p output voltage. Moreover, the power consumption with no load is maintained within 350 mu W.